Calculations
Area
Chromatogram area is calculated only where peaks are, in particular where:
-
Detected peaks, that are within Integrable Range. These regions are highlighted green when you’re editing Integration Method. If such a peak touches non-integrable ranges, then such peak isn’t considered.
-
Manually selected or modified peaks. Even if it’s inside the non-integrable range, the area of such peak gets included in the chromatogram area.
The actual integration uses Peaksel uses the Trapezoidal rule.
Peak Retention Time
Retention Time (RT) is taken as the highest point of a peak.
However, Peaksel will never use the boundary point as RT, so if the highest point is at the peak boundary (happens in case of round peaks and shoulders), then we neighbor point is used.
Peak Base, BPM, λmax
If a detector measures spectra (Mass Spec, DAD/PDA), then for every peak we calculate its mean spectrum (is available if you double-click on the peak label). The base is the highest point of in the mean spectrum.
The Base of the peak corresponds to:
-
Base Peak Mass in case of Mass Spectrometry
-
λmax (the most absorbing wavelength) in case of DAD/PDA